589-19-3-7626-0142, EK, CT DOSE PROFILER


Kansas, United States
Government : Federal
RFP
Go to the link
This document has expired, therefore the above link may no longer work.

Statement of Work:
EKHCS, HTM Department
The HTM Department desires to replace its obsolete Fluke TNT 12000 Diagnostic X-ray QA system (589-EE138200) with Name Brand or Equal to the BLACK PIRANHA Model 557. This system can be used to test radiation-producing equipment and other radiological modalities for compliance within certain predetermined, accepted metrics and determine both safety and efficacy of using these devices on patients. This replacement system shall be upgradable to meet anticipated technology advancements over the next seven years. The system can be used to keep the facility in compliance with requirements outlined in Joint Commission (JC) Environment of Care (EC) 02.04.03 Elements of Performance (EP) 21, 22, 23, 24 and 34.
The system shall be: Name Brand or Equal to the BLACK PIRANHA Model 557

Transmitting all invasive and non-invasive test data (mA, mAs, kVp, Time, and Dose) via Bluetooth wireless
Vendor shall produce Federal Information Processing Standards (FIPS) 140-2 certificate number, is certified
Ability to check the detector position
Solid state detector
Base unit (BP) includes ability to measure all basic parameters.
Probes shall operate on battery for at least 12 hours continuous operation on a full charge.
Being calibrated a minimum of every two (2) years
Calibration must be meet all NIST standards.
Downloading data in an MS Excel spreadsheet format both in real time and as a completed QA report. Alternatively, the system should be able to save test data directly into a Word or Excel file and transfer completed chart to a remote document.
Retaining 15-hour battery life
Testing RF, CT, and Dental Modalities
Vendor must provide training to two individuals on-site.
Vendor should provide at a minimum a two year manufactures warranty and have an option to purchase an extended warranty.
Any software required to display or manipulate data must be included. Software should function in a Windows 7 or 10 Operating System (OS).