COMBINED SYNOPSIS SOLICITATION
(i) This is a combined synopsis/solicitation for Phillips Cardiac Monitors, BRAND NAME , as prepared in accordance with the format in Subpart 12.6, as supplemented with additional information included in this notice. This announcement constitutes the only solicitation; quotes are being requested and a written solicitation will not be issued. A firm-fixed price purchase order is anticipated.
(ii) The solicitation number is 36C24418Q0333 and is issued as a request for quotation (RFQ).
(iii) The solicitation document and incorporated provisions and clauses are those in effect through Federal Acquisition Circular 2005-95
(iv) This solicitation is set aside 100% for Service-Disabled Veteran-Owned Small Businesses and the associated NAICS 339113 code has a small business size standard of 500.
(v) Contract Line Items (CLIN):
ITEM NUMBER
DESCRIPTION OF SUPPLIES/SERVICES
QUANTITY
UNIT
UNIT PRICE
AMOUNT
0001
SH coverage up to 10,000 sq ft 1.4GHz Smart hopping coverage per square foot (up to 10,000 sq ft). Includes all smart hopping materials necessary to achieve the required wireless coverage within the
LOCAL STOCK NUMBER: MXU0480

5,500.00
EA
__________________
__________________
0002
Smart hopping - Enhanced implementation 1.4GHz Smart hopping infrastructure implementation per square foot. Includes all labor necessary to implement the required wireless coverage within the defined
LOCAL STOCK NUMBER: MXU0475

5,500.00
EA
__________________
__________________
0003
Installation Site Services A04 Facilities Implem Solutions
LOCAL STOCK NUMBER: H1028B 890500

1.00
EA
__________________
__________________
0004
Installation Site Services A07 Project Mngemt Impl Solutions
LOCAL STOCK NUMBER: H1028B980500

1.00
EA
__________________
__________________
0005
Installation Site Services A17 US Government Facilities Service
LOCAL STOCK NUMBER: H1028B980500

1.00
EA
__________________
__________________
0006
Remote Multi-Video Solution Sender 8
LOCAL STOCK NUMBER: 989805700066

1.00
EA
__________________
__________________
0007
Remote Multi-Video Solution Receiver
LOCAL STOCK NUMBER: 989805700067

6.00
EA
__________________
__________________
0008
Cat5/5e/6 Cable Terminate & TIA/EAI Test
LOCAL STOCK NUMBER: MCU0472

193.00
EA
__________________
__________________
0009
Smart hopping - Enhanced implementation 1.4GHz Smart hopping infrastructure implementation per square foot. Includes all labor necessary to implement the required wireless coverage within the defined
LOCAL STOCK NUMBER: MCU0475

21,583.00
EA
__________________
__________________
0010
Biomedical Training Courses A17 CMS2024 PIIC iX Basic Service Trng
LOCAL STOCK NUMBER: MXU0011

2.00
EA
__________________
__________________
0011
Biomedical Training Courses A21 CMS2032 MX40 ITS/Cableless PIIC iX
LOCAL STOCK NUMBER: MXU0011

2.00
EA
__________________
__________________




GRAND TOTAL
__________________

ITEM NUMBER
QUANTITY
DELIVERY DATE
0001
SHIP TO:
646 UD DEPARTMENT OF VETERANS AFFAIRS
VA PITTSBURGH HEALTHCARE SYSTEM/2A123
UNIVERSITY DRIVE C
PITTSBURGH, PA 15240
USA

ALL ITMES
DAYS ARO


(vi) Comparable products must be Brand Name in the following specifications:
DEPARTMENT OF VETERAN AFFAIRS
STATEMENT OF WORK
Cardiac Monitors


GENERAL INFORMATION

Scope of Work: The contractor shall provide all resources necessary to accomplish the deliverables described in this statement of work (SOW), except as may otherwise be specified. Procurement is for the replacement of the cables which are needed to install new Philips cardiac and hemodynamic monitoring system (hereafter referred to as cardiac monitoring system or cardiac monitors) in the Emergency Department (ED), the Critical Care Units, and the 9 West Observation Unit at the VA Pittsburgh, University Drive Division Pittsburgh PA 15240.

Background: VA Pittsburgh purchased new Philips Cardiac monitors in FY17. After a review of the current cabling in these units, Philips provided a detailed report showing not all current cabling met the requirements for their system. As a result, VA Pittsburgh would like to remove all old cabling and replace with new cabling to ensure proper installation and operation which is needed for patient care.

B. Cabling/Network
Turn-Key Installation providing all hardware and accessories for the Philips cardiac monitoring system, services shall include removal of current and older patient monitoring cabling/hardware.
This is a standalone network just for cardiac monitors
All cabling and accessories to provide a complete and functional system
All current network hardware: switches, routers, hubs, etc. must be replaced
Unit to unit connection charge shall include all cable runs, hubs, switches, routers and any other hardware or software required.
All above ceiling cabling runs shall be tie-wrapped and placed in telephone/data trough, in a conduit, or properly routed through interstitial area per hospital facilities requirements and local electric codes. This includes facility approval and post installation inspection when running cable through a fire barrier or fire wall.
Cables shall be bundled neatly, labeled, and in an orderly manner especially when cables converge with facility network hardware.
Cables shall be marked at each end indicating the termination point of the other end.
All new network cabling, terminations, and any patch panels used shall be CAT5E/CAT6 certified.
All cables shall be terminated TIA568B.
Any cable run through plenum space shall be plenum rated according to NEC and applicable fire codes.
All cable runs shall be tested and certified in accordance with TSB-67 and TIA/EIA 568-B or latest TIA/EIA Revisions.
The offeror shall provide a copy of all test results to the COR in electronic format that can be displayed and/or viewed.
Cable length shall be included in this report.
Documentation of the network shall also be provided and include a marked up drawing (as built) showing jacks and room locations.
The system will be configured to view patients throughout the hospital or care areas.
Drawings that indicate the location of the monitoring devices will be provided for the Medical Center.
Interface will be compliant with VA National interface standard.
Any offsite server or network maintenance or support provided by the offeror can only be done through VPN access after the offeror has obtained such access from the VA.
All networking hardware shall be rack mounted in room designated by Facility Project Manager or COR.
All installed monitors and hallway monitors will have supportive backing to improve weight distribution of the monitor and mounting arm.
All installed monitors will have power run through conduit.
Telemetry receivers will adhere to NFPA99 and NFPA101, especially relating to distance from sprinkler heads. Vendor may be required to install.
Offeror will adhere to ICRA Level 2 guidelines during installation.
Offeror agrees to mutually agreed upon timeframes for project completion and go-live.
Offeror will provide, or subcontract, mounting brackets to attach to existing GCX mounting arms.
Cabling and telemetry field will be needed.
Vendor to provide ICRA if applicable which is to be coordinated with biomedical engineering and COR.

C. HL7 Solution (If Applicable)
Offeror shall provide an HL7 interface solution allowing the exchange of data from point-of-care equipment through HL7 messaging to VISN 4 s Clinical Information System (CIS) / Anesthesia Record Keeper (ARK).

Included in the quote the offeror shall have a separate line item for 1 (one) year HL7 software support.

D. Installation and Implementation Services

The offerors shall provide coordinated professional installation and implementation project management services to implement the system specified in the scope of work at the Pittsburgh VA Healthcare System. Installation meetings and related activities shall begin within one week of contract award. This service shall include removal of existing system and turnkey installation. Offerors shall provide a firm estimate of working days required from date of purchase order to go-live based on scope of work.

The price quote shall include cost of installation which consists of assembling, positioning, and mounting of all equipment listed on the delivery order and connections of all cables. The hospital is responsible for furnishing all conduit and raceways unless specified otherwise on the delivery order. The equipment contractor is responsible for furnishing and pulling interconnecting wiring and cabling through conduit and raceways, and for making any connections. Interconnecting wiring and cabling which do not run through conduit and raceways shall be furnished and installed by the equipment contractor. It is the responsibility of the equipment contractor to install junction boxes, wall/ceiling mounts, and support structures supplied by the equipment contractor. The equipment contractor must provide well qualified field engineers or technicians to install and conduct all necessary tests which shall begin within (10) ten days after receipt of notice to proceed from the Contracting Officers Representative (COR) at the affected hospital.
Offeror will negotiate the work schedule with the COR assigned to the project. Compliance with this requirement shall be manifest by the continuous presence of the engineers or technicians on the job site during the daily working period. The contractor must provide the physical movement of the equipment from the storage point at final destination, to the area of installation, and the uncrating of the equipment. Vendor to provide ICRA if applicable which is to be coordinated with Biomed and COR

Rigging and special handling costs, if required, to move the equipment from dock area to the installation site within the consignee's premises, shall be borne by the equipment contractor.

Upon receipt of notice to proceed with installation, it shall be the contractor's responsibility to inform the Contracting Officer of any problems which may be anticipated in connection with installation or which will affect optimum performance once installation is completed. Such matters as inadequacy of power supply, limitations of site or inadequate preparation of site shall be reported prior to start of installation. Installation shall not proceed under such circumstances until authorized by the Contracting Officer.

In the event that progress of the installation is interrupted through no fault of the contractor, the continuous installation referenced in the preceding paragraphs may be terminated until such time as the cause of delay has been eliminated, and then shall be resumed within twenty-four (24) hours after the contractor has been notified that work may again proceed. Such termination of continuous installation shall be made only after two (2) hours notice has been given to the Contracting Officers Representative (COR) or person acting in that capacity at the hospital receiving installation. Contractor must notify the Contracting Officer within 48 hours of termination of installation.

Unless otherwise authorized by the Contracting officer, installation of equipment shall in no event exceed sixty (60) days from the date of notice to proceed with installation.
Offers shall include new equipment and infrastructure for all care areas. Existing equipment will not be upgraded or utilized with the new system, with the exception of cabling/antennas.

E. IHE INTEGRATION STATEMENT

Vendor shall provide IHE (Integrating the Healthcare Enterprise) integration statement(s) (IIS) for proposed solution. (See http://www.ihe.net/ for more information)
F. MAINTENANCE REQUIREMENTS

Licensing Software updates: The offeror shall provide, at no charge for the life of the monitoring system, all computer software, access keys or codes, or external devices required for the operation, calibration, or repair of the equipment purchased. Any such items not listed on the price quote and required for maintenance of the system, shall be taken as included with the purchase of the system. Any upgrades or changes to the maintenance software, hardware, or access keys or codes shall be provided at no charge to the medical center during the time the equipment is operational at this facility. All application software licenses are included in the purchase of the equipment and shall not a require renewal charge for the period of time the equipment is in use in the facility.

INSTALLATION - The Contractor shall provide a system Installation Plan (IP) to the Contracting Officer Representative (COR) for approval prior to the start of work. The COR shall approve the IP and return the contractor. The Installation Plan shall include estimated timeline of start and completion and estimate unexpected delays/lead time.

Perform staging, programming, and configuration of all system infrastructures.

Perform all necessary test/audits on all system infrastructures.

TEST AND ACCEPTANCE: Contractor shall test all equipment after installation. The government shall accept equipment once installation and successful testing has been completed and approved.

WARRANTY/SERVICE CONTRACT: All equipment and materials shall come with a standard one year warranty. Warranty shall begin after installation of equipment and completion of tests.

G. CONTRACTOR RESPONSIBILITIES: The contractor shall be responsible for the following:
Contractor shall provide qualified personnel to perform install of the cardiac monitoring system expansion.

Contractor shall be responsible for all system/software testing after installation and prior to government acceptance.

Contractor is responsible for ensuring the proper disposal of all debris generated from installation activities.

Contractor is responsible for securing all materials, equipment and tools while on government property or in government facility. Government is not liable for any lost or stolen items that are not properly secured.

Vendor to provide ICRA class II if applicable which is to be coordinated with Biomed and COR
Provides active means to prevent air-borne dust from dispensing into atmosphere
Water mist work surfaces to control dust while cutting
Seal unused doors with duct tape
Block off and seal air vents
Wipe surfaces with disinfectant
Contain construction waste before transport in lightly covered containers
Wet mop and /or vacuum with HEPA filtered vacuum before leaving the area
Place duct mat at entrance and exit of work area
Remove or isolate HVAC system in areas where work is being performed.


H. GOVERNMENT RESPONSIBILITIES: The Government shall provide the following:

Provide site access and escorts through work areas as needed.
Provide primary electrical power.
Provide adequate space for the system equipment


I. OTHER CONSIDERATIONS:

CONTRACTING OFFICER REPRESENTATIVE (COR): The Contracting Officer Representative shall be the technical point of contact for all work related requirements. The COR does not have authorization to change, alter, or remove any requirements stated in the performance work statement. The Contracting Officer is the only government authorized person to change the performance work statement or any resultant contract.

PLACE OF PERFORMANCE: Installation shall occur at the Emergency Department (ED) and the Surgical Intensive Care Units (SICU) VA Pittsburgh, University Drive Division, Pittsburgh, Pennsylvania 15240.

WORK HOURS: Operational Hours are 0730-1600, Monday-Friday

DELIVERY: All deliveries shall be coordinate with the COR before arrival. Failure to coordinate delivery will be grounds to refuse delivery.

(viii) The provision at 52.212-1, Instructions to Offerors -- Commercial, applies to this acquisition and the following clauses AND instructions are added as addenda:
CLAUSES:
52.214-21 Descriptive Literature (APR 2002)
852-219-10 VA Notice of Total Service-Disabled Veteran-Owned Small Business Set-Aside

(a) Gray market items are Original Equipment Manufacturer s (OEM) goods sold through unauthorized channels in direct competition with authorized distributors. This procurement is for new OEM medical supplies, medical equipment and/or services contracts for maintenance of medical equipment (i.e. replacement parts) for VA Medical Centers.B No remanufactures or gray market items will be acceptable.
B
(b) Vendor shall be an OEM, authorized dealer, authorized distributor or authorized reseller for the proposed medical supplies, medical equipment and/or services contracts for maintenance of medical equipment (i.e. replacement parts), verified by an authorization letter or other documents from the OEM, such that the OEM s warranty and service are provided and maintained by the OEM.B All software licensing, warranty and service associated with the medical supplies, medical equipment and/or services contracts for maintenance of medical equipment shall be in accordance with the OEM terms and conditions.

(c) The delivery of gray market items to the VA in the fulfillment of an order/award constitutes a breach of contract.B Accordingly, the VA reserves the right enforce any of its contractual remedies. This includes termination of the contract or, solely at the VA s election, allowing the Vendor toB replace, at no cost to the Government, any remanufactured or gray market item(s) delivered to a VA medical facility upon discovery of such items.B
(ix) Evaluation of this requirement will be based on PRICE ONLY.
(x) Offerors are advised to include a completed copy of the provision at 52.212-3, Offeror Representations and Certifications -- Commercial Items, with its offer if has not been completed on SAM.gov.
(xi) The clause at 52.212-4, Contract Terms and Conditions -- Commercial Items, applies to this acquisition and the following clauses are added as addenda:
(End of Clause)
52.252-2 CLAUSES INCORPORATED BY REFERENCE (FEB 1998)
This contract incorporates one or more clauses by reference, with the same force and effect as if they were given in full text. Upon request, the Contracting Officer will make their full text available. Also, the full text of a clause may be accessed electronically at this/these address(es):
http://farsite.hill.af.mil
https://acquisition.gov/far
(End of Clause)

852.203-70 Commercial Advertising (JAN 2008)
852.246-71 Inspection (Jan 2008)
(xii) The clause at 52.212-5, Contract Terms and Conditions Required To Implement Statutes Or Executive Orders -- Commercial Items, applies to this acquisition and the following additional FAR clauses cited in the clause are applicable to the acquisition:
52.204-10 Reporting Executive Compensation & First-Tier Subcontract Awards (OCT 2015)
52.209-6 Protecting the Government s Interest When Subcontracting with Contractors Debarred, Suspended, or Proposed for Debarment (OCT 2015)
52.219-28 Post Award Small Business Program Representation (JUL 2013)
52.222-19 Child Labor--Cooperation with Authorities and Remedies (FEB 2016) (E.O. 13126)
52.222-21 Prohibition of Segregated Facilities (APR 2015)
52.222-26 Equal Opportunity (APR 2015)
52.222-36 Equal Opportunity for Workers with Disabilities (JUL 2014)
52.222-50 Combating Trafficking in Persons (MAR 2015)
52.223-18 Encouraging Contractor Policies to Ban Text Messaging While Driving (AUG 2011)
52.225-3 Buy American--Free Trade Agreements--Israeli Trade Act (MAY 2014)
52.225-13 Restrictions on Certain Foreign Purchases (JUNE 2008)
52.232-34 Payment by Electronic Funds Transfer--Other than System for Award Management (JUL 2013)
(xiii) There are no additional contract requirements, terms or conditions.
(xiv) The Defense Priorities and Allocations System (DPAS) ratings are NOT APPLICABLE.
(xv) Quotes must be emailed to amanda.saunders@va.gov and received no later than NOON EST on 2/23/2018 BY NOON EST
Quotes may be submitted on this document (If you use this form, Include your company information name, address, phone number, POC, POC email, and DUNS #) or the vendor s own form.
If you are not the manufacturer than an AUTHORIZED DISTRIBUTOR LETTER needs to be provided with your quote in order to be considered for the award.
NO LATES WILL BE ACCEPTED
(xvi) For information regarding the solicitation, please contact Amanda Saunders at amanda.saunders@va.gov

AMANDA SAUNDERS
AMANDA.SAUNDERS@VA.GOV

amanda.saunders@va.gov