New Freestanding steam autoclave with an approximate interior chamber size of 20H x 20W x 38D and stainless steel exterior enclosure. Steam supplied by three phase electric integral steam generator (208V). At least gravity and prevacuum cycles standard. Preference given to units with water saving technology to reduce water usage while idle to approximately 100 gallons per day. At least four programmable cycles and an impact printer. Radial arm door hinged on the left hand side Two interior shelves. If remanufactured the unit must have new solenoid valves, a new wiring harness and all new piping and a hydrostatically tested vessel.